5.2
Impact Factor
Generic selectors
Exact matches only
Search in title
Search in content
Post Type Selectors
Search in posts
Search in pages
Filter by Categories
Corrigendum
Current Issue
Editorial
Erratum
Full Length Article
Full lenth article
Letter to Editor
Original Article
Research article
Retraction
Retraction notice
Review
Review Article
SPECIAL ISSUE: ENVIRONMENTAL CHEMISTRY
5.3
Impact Factor
Generic selectors
Exact matches only
Search in title
Search in content
Post Type Selectors
Search in posts
Search in pages
Filter by Categories
Corrigendum
Current Issue
Editorial
Erratum
Full Length Article
Full lenth article
Letter to Editor
Original Article
Research article
Retraction
Retraction notice
Review
Review Article
SPECIAL ISSUE: ENVIRONMENTAL CHEMISTRY
View/Download PDF

Translate this page into:

Original article
12 (
7
); 1707-1714
doi:
10.1016/j.arabjc.2014.08.024

Development and validation of a UPLC method for quantification of antiviral agent, Acyclovir in lipid-based formulations

Kayyali Chair for Pharmaceutical Industries, College of Pharmacy, King Saud University, Riyadh, Saudi Arabia

⁎Corresponding author at: Kayyali Chair for Pharmaceutical Industries, Department of Pharmaceutics, College of Pharmacy, P.O. Box 2457, King Saud University, Riyadh 11451, Saudi Arabia. Tel.: +966 (1) 4677372; fax: +966 (1) 4676295. mkazi@ksu.edu.sa (Kazi Mohsin)

Disclaimer:
This article was originally published by Elsevier and was migrated to Scientific Scholar after the change of Publisher.

Peer review under responsibility of King Saud University.

Abstract

Purpose

The objective of the current study is to evaluate the Ultra Performance Liquid Chromatography (UPLC) method for quantification of Acyclovir in lipid-based formulations.

Method

A simple, rapid, reliable and precise reversed phase UPLC method has been developed and validated according to the regulatory guidelines, which composed of isocratic mobile phase; 0.25% formic acid (FA) in Milli-Q water with a flow rate of 0.5 ml/min, and column BEH C18 (2.1 × 50 mm, 1.7 μm). The detection was carried out at 254 nm.

Results

The developed UPLC method was found to be rapid (1.2 min run time), selective with well resoluted Acyclovir peak (0.89 min) from different lipid matrices and sensitive (Limit of Detection (LOD) was 0.3 ppm and Lower Limit of Quantification (LLOQ) was 1 ppm). The accuracy and precision were determined and were perfectly matching with the standard FDA limits.

Conclusion

The study showed that the proposed UPLC method can be used for the assessment of drug purity, stability, solubility and lipid-formulation release profile with no interference of excipients or related substances of active pharmaceutical ingredient.

Keywords

Acyclovir
UPLC
Lipid based formulation
Validation
1

1 Introduction

Acyclovir (ACV) is chemically known as 2-amino-1,9-dihydro-9-[(2-hydroxyethoxy)methyl]-6H-purin-6-one (Fig. 1). Acyclovir is the most commonly used oral antiviral agent for the treatment and prophylaxis of initial and recurrent episodes of genital and labial herpes and also for the acute treatment of herpes zoster and the varicella (chickenpox) in immunocompetent individuals. As an antiviral therapy it is extremely selective and low in cytotoxicity. According to the pharmacokinetics data ACV is poorly water soluble and it has low, variable and incomplete absorption which leads to poor bioavailability ranging from 10 to 30%. Maximum plasma concentration is reached within 1.5–2.5 h, when ACV is administered orally (Arnal et al., 2008).

Chemical structure of Acyclovir (MW: 225.21, pKa: 2.27 and 9.25).
Figure 1 Chemical structure of Acyclovir (MW: 225.21, pKa: 2.27 and 9.25).

ACV was previously analyzed with several analytical methods like High Performance Liquid Chromatography (HPLC) (Teshima et al., 2003, Basavaiah et al., 2003), micellar liquid chromatography (Peh and Yuen, 1997; Macka et al., 1993), gas chromatography (Leis et al., 2002, Rakestraw, 1993), capillary electrophoresis (Vo et al., 2002, Reichova et al., 2002), radioimmunoassay (Tadepalli and Quinn, 1996), and potentiometry (Abdel-Ghani et al., 2002). None of these methods were specially developed in lipid based formulations except biological fluids and few in Acyclovir dosage forms (suspension). There are many drawbacks with the previously mentioned analytical methods such as lengthy and tedious process, lack of required sensitivity or may have required synthetic/special reagents or detectors. Thus, an improved analytical method for ACV should be developed demanding more sensitivity, speed and also required to meet the stability-indicating parameter.

In the current project, the assay method for ACV was developed and validated using a relatively new Ultra Performance Liquid Chromatography (UPLC) system that carries many advantages over HPLC or any other conventional analysis method which reduces analysis time and solvent use significantly. The solvent used as the mobile phase was only water mixed with 0.25% formic acid. Within the experimental method development, it was also needed to carry out stability studies under forced acidic, alkaline, thermal, and oxidative degradation processes. The special feature of UPLC device allows for the system to withstand high back pressure without any harmful effect to the analytical column or the whole device (Abdel-Hamid et al., 2012, Nováková et al., 2006). This proposed method was also successfully applied to the analysis of lipid-based formulations containing ACV with no interference from dosage form excipients. The method was validated with respect to the standard FDA guidelines (US Dept. of Health and Human Services, 2001).

2

2 Materials

ACV (purity >99.5) was generously donated by Riyadh Pharma Medical and Cosmetic Products Co. LTD, Riyadh 11451, Saudi Arabia. Imwitor 308 (I308) was kindly supplied by Sasol Germany GmbH (Werk Witten, Witten-Germany). Propylene glycol (PG) and Tween 80 (T80) were obtained from Winlab, BDH and Merck-Schuchardt. The mobile phase was HPLC-grade 0.25% formic acid which was prepared by adding 5 ml of formic acid (obtained from BDH laboratory, BDH Chemicals Ltd., Poole, UK with purity of 98–100%) to two liters of Milli-Q water. The high purity Milli-Q water was obtained through a Milli-Q Integral Water Purification System (Millipore, Bedford, MA). All other reagents were of analytical grade and used without further purification.

3

3 Methods

3.1

3.1 UPLC chromatographic conditions

Chromatographic separation was developed and optimized with respect to the stationary and mobile phase compositions, flow-rate, sample volume, and detection wavelength. The study employed a highly sensitive UPLC system that consisted of an Acquity® UPLC binary solvent manager equipped with an Acquity® automatic sample manager and a Photodiode Array (PDA) eλ detector obtained from Waters (Waters Inc., Bedford, MA, USA). Separation was achieved by reverse-phase isocratic elution using a mobile phase consisting of 0.25% formic acid (0.25% FA) in Milli-Q water delivered at a flow rate of 0.5 ml/min through an Acquity® UPLC BEH C18 column (2.1 × 50 mm, 1.7 μm) kept at 50 °C. The run time was 1.2 min. Freshly prepared mobile phase was filtered through an online 0.20 μm filter and degassed continuously by an online degasser within the UPLC system. The detector wavelength was set at 254 nm and the injection volume was 1.0 μl.

3.2

3.2 Preparation of stock solution, calibration standards and QC samples

Stock solution of ACV was prepared by dissolving 50 mg ACV powder in 50 ml distilled water, resulting in a solution containing 1000 ppm ACV. Serial standard dilutions (9 points) of ACV were prepared to cover the concentration range of 1–400 ppm. These standard solutions were stored at 4 °C before use. Calibration curves were obtained by plotting peak area against standard drug concentration and regression equations were computed thereby. Four quality control (QC) samples with the selected concentrations (1.48, 64.48, 287.76, and 377.12 ppm) were prepared to cover the desired range. QC samples were prepared by spiking the lipid-formulation with known amount of ACV, then diluting the mixture with appropriate volume of water containing 0.25% FA.

3.3

3.3 Formulation matrix effect

Direct spectrophotometric analysis of the drug compound may experience spectral overlapping between the analyte and excipients used in the dosage form, which significantly affects the sensitivity, accuracy and precision of the method (Yu et al., 2008). Therefore it was required to perform UPLC spectral scanning for both drug-containing and drug-free formulations to detect any possible interference. In this study, the analyte (ACV) was loaded in lipid based formulation from low to high concentrations in order to test the extent of the matrix effect.

3.4

3.4 Bioanalytical method validation

The above mentioned developed method was validated in terms of linearity, specificity, precision and accuracy, Limit of Detection (LOD) and Lower Limit of Quantification (LLOQ) according to the standard guidelines of bioanalytical method validation (US Dept. of Health and Human Services, 2001; Gao et al., 2011) by FDA.

3.4.1

3.4.1 Linearity and range

Appropriate volumes of ACV stock solution (1000 ppm) were utilized to prepare nine non-zero standard drug concentrations covering the calibration range of 1–400 ppm. Four different QC samples were prepared by spiking known concentrations of ACV within the detection range of 1–400 ppm.

Each standard solution (1, 2, 5, 10, 20, 50, 100, 200 and 400 ppm) has been injected six consecutive times daily on three consecutive days for validation. Calibration solutions were injected in ascending order in each validation run and the other samples were distributed randomly through the run.

Linear regression equation and correlation coefficient (R2) were employed to statistically evaluate the linearity of the results (Ali et al., 2006; Lister, 2005; Al-Hadiya et al., 2010).

3.4.2

3.4.2 Specificity

Specificity of the method needed to be assessed for the evaluation of the matrix effect between the drug and different lipid-based formulations (details in matrix effect section).

To evaluate the specificity of the method, drug free lipid formulation (QCzero) samples were carried out through the assay procedure and the retention times of the lipid formulation components were compared with those of ACV analyte (Al-Hadiya et al., 2010). In addition, specificity of the method toward the intact drug was also studied by determination of the resolution (R) between the drug peak and the nearest degradation product if present (Dongre et al., 2008).

3.4.3

3.4.3 Accuracy and precision

Intra-day accuracy and precision evaluations were performed using six replicate determinations of nine ACV standards within the same day. Inter-day accuracy and precision were assessed by six replicate analyses of the following: LLOQ, low, medium and high QC samples on three consecutive days. The overall precision of the method was expressed as relative standard deviation and accuracy of the method was expressed in terms of % drug recovered (Gao et al., 2011).

3.4.4

3.4.4 Limit of Detection (LOD) and Lower Limit of Quantification (LLOQ)

The LOD and LLOQ were determined by serial dilutions of ACV stock solutions in order to obtain signal to noise (S/N) ratio of at least ≈3:1 for LOD and ≈10:1 for LLOQ (Lister, 2005).

4

4 Results and discussion

4.1

4.1 Chromatographic separation

Fig. 2 shows the representative chromatograms of blank sample (A), standard solution of ACV (B), drug-free lipid formulation (QCzero) sample (C), drug-containing lipid formulation (QC2) sample (D).

UPLC chromatograms of blank sample (A), standard solution of ACV at concentration of 400 ppm (B), drug-free lipid formulation (QCzero) sample (C), and drug-containing lipid formulation (QC2) sample (D).
Figure 2 UPLC chromatograms of blank sample (A), standard solution of ACV at concentration of 400 ppm (B), drug-free lipid formulation (QCzero) sample (C), and drug-containing lipid formulation (QC2) sample (D).

Separation and detection of ACV without any interference were ideal by the developed UPLC assay. The chromatographic result of UPLC technique in the current study shows that the sensitivity and selectivity of this procedure are good enough to determine ACV within the available excipients. The ACV analyte was well separated at a retention time of ∼0.89 min without any interference from the excipients (Fig. 2A–D). The total run time was ∼1.2 min, where the peaks were of good shape and completely resolved.

4.2

4.2 Matrix effect

The assessments of matrix effect represented an integral part of validation for quantitative analysis of drug in lipid based formulations within this analytical method development. The matrix effects at four concentration levels of ACV were assessed. In the present analysis, the overall process efficiency demonstrated that the proposed method was practically free from relative matrix effects for the determination of ACV in lipid based formulation.

4.3

4.3 Forced degradation study

The force degradation experiment in the studies was conducted by treating model drug ACV with 1 N HCl, 1 N NaOH, 3% H2O2 under 100 °C temperature for 2 h at water bath. In case of thermal degradation the experiment was carried out under 120 °C temperature in an oven. All the ACV sample solutions used in forced degradation studies were prepared with final concentrations of 400 μg/ml (400 ppm). The results of the degradation study showed that the concentrations of ACV in the samples were lowered compared to the original concentrations. This degradation study was particularly important for the analysis of lipid based formulation, which was not performed previously (Basavaiah et al., 2003, Macka et al., 1993).

4.3.1

4.3.1 Acid degradation

This study has been made to show the degradation products of ACV using the acid degradation method. ACV stock solution of 400 ppm was prepared; 10 ml of it was transferred to a 100 ml volumetric flask and then10 ml of 1 M hydrochloric acid solution was added. The solutions in the flask were mixed well and kept for 2 h at 100 °C temperature. After cooling the solution at room temperature, 10 ml of 1 M sodium hydroxide solution was added to a 100 ml volumetric flask to neutralize and then diluted to 100 ml with the mobile phase, 0.25% FA in water. The chromatogram of the acid degradation is shown in Fig. 3A.

Typical UPLC chromatograms of: (A) acid hydrolysis-degraded ACV, (B) base hydrolysis-degraded ACV, (C) oxidative-degraded ACV, and (D) thermal-degraded ACV.
Figure 3 Typical UPLC chromatograms of: (A) acid hydrolysis-degraded ACV, (B) base hydrolysis-degraded ACV, (C) oxidative-degraded ACV, and (D) thermal-degraded ACV.

4.3.2

4.3.2 Basic degradation

10 ml of the stock solution (400 ppm) was transferred into a 100 ml volumetric flask and treated with 10 ml of 1 M sodium hydroxide solution. The contents of the flask were mixed well and kept for 2 h at 100 °C temperature. After cooling the solution at room temperature, 10 ml of 1 M hydrochloric acid solution was added to a 100 ml volumetric flask to neutralize and then diluted to 100 ml with the mobile phase, 0.25% FA in water. The result of the basic degradation is shown in Fig. 3B.

4.3.3

4.3.3 Oxidation

10 ml of the stock solution (400 ppm) was transferred to a 100 ml volumetric flask and treated with 10 ml of 3% hydrogen peroxide H2O2 and then heated in water bath at 100 °C for 2 h. The sample was cooled to room temperature and then the volume completed to 100 ml with the mobile phase 0.25% FA in water. The sample was injected in the UPLC system. The chromatogram of the oxidation data is shown in Fig. 3C.

4.3.4

4.3.4 Thermal degradation

Thermal degradation experiment was conducted at elevated temperature. 10 ml of the stock solution (400 ppm) was transferred to a 100 ml volumetric flask, and kept in oven at 120 °C for 24 h. After cooling the solution at room temperature, the 0.25% FA in water was added up to 100 ml as diluent. The chromatogram of the thermal degradation study is shown in Fig. 3D.

4.4

4.4 System suitability

The performance parameters of the developed UPLC system method were determined by analyzing standard working solution (400 ppm). Analytical parameters such as capacity factor (k) and number of theoretical plates (N) were calculated against the availability of ACV peak only. In addition, system repeatability was determined by calculating the relative standard deviation (% RSD) of the peak areas of six consecutive injections of working standard solution (400 ppm). From the overall analysis, the variation (% RSD) in the peak area (0.47%), capacity factor (k = 1.66) and number of theoretical plates (N = 5069.4) results shows the acceptable performance of the systems (Table 3).

4.4.1

4.4.1 Bioanalytical method validation

4.4.1.1
4.4.1.1 Linearity and range

The peak area response of ACV was linear over the concentration range between 1 and 400 ppm (Fig. 4). The results of linear regression give the following mean equation: y = 113.906 ( ± 1.30946 ) x + 21.62037 ( ± 2.119667 ) where y denotes the peak area of the analyte and x denotes the concentration of the analyte. These results show excellent linearity over the interval studied with correlation coefficient(r) = 0.9993 (±0.007272) (Mulholland and Hibbert, 1997).

UPLC calibration curve of ACV in water containing 0.25% formic acid (FA).
Figure 4 UPLC calibration curve of ACV in water containing 0.25% formic acid (FA).

4.4.1.2
4.4.1.2 Accuracy recovery

The accuracy was calculated as the % of drug recovered after analysis relative to the corresponding nominal concentrations. The intra-day (Table 1) accuracy (mean recoveries, mean ± SD) was found between 90% and 106.33% and the inter-day (Table 2) accuracy (mean recoveries, mean ± SD) was found between 96.09% and 107.04%. The results from the drug recovery suggest that the accuracy of the assay method has been well matched with the acceptance criteria of FDA guidelines (US Dept. of Health and Human Services, 2001).

Table 1 UPLC data of intra-day back-calculated ACV concentrations of the calibration standards in water containing 0.25% formic acid.
Nominal conc. (ppm) Back calculated concentrations (ppm)
1st 2nd 3rd 4th 5th 6th Mean SD Precision % Accuracy %
1 0.9 0.9 0.9 0.9 0.9 0.9 0.9 0 0 90
2 2.1 2.1 2 2 2.1 2 2.05 0.055 2.67 102.50
5 5.3 5.4 5.3 5.2 5.4 5.3 5.32 0.052 0.97 106.33
10 10.6 10.4 10.2 10.4 10.4 10.5 10.42 0.160 1.54 104.17
20 20 19.6 19.5 18 18 18.1 18.87 0.845 4.49 94.33
50 49.9 46.1 49.9 49.8 49.9 50.4 49.33 1.604 3.25 98.67
100 99.8 99.8 99.3 98.4 100.3 99.5 99.52 0.383 0.38 99.52
200 200.4 200.3 202.3 201.3 202.8 202.1 201.53 1.071 0.53 100.77
400 403.6 400 399.3 399.6 398.1 399.8 400.07 1.876 0.47 100.02
Table 2 UPLC data of inter-day accuracy (recovery) and precision of ACV QC samples.
Day of analysis QC1 QC2 QC3 QC4
1.48 ppm 64.48 ppm 287.76 ppm 377.12 ppm
1st day 1.4 66.5 289.8 377.8
1.3 66.9 289.4 377.6
1.3 66.6 291 373.1
1.3 66.6 290.2 376.1
1.3 60.8 290.4 373.8
1.3 66.7 289.4 373.9
2nd day 1.4 69.7 290.3 376.5
1.5 69.7 290.9 375.1
1.4 70.2 287.9 374.7
1.4 69.8 289.3 376.1
1.5 69.7 290.2 376.2
1.5 70.1 289.3 374.5
3rd day 1.5 71.4 292.6 376.6
1.5 71.5 290.3 376.4
1.5 71.8 290.4 376.1
1.5 71.5 293.8 375.9
1.5 71.7 290.3 378.9
1.5 71.2 292.2 375.8
Mean 1.42 69.02 290.43 375.84
SD 0.09 2.84 1.36 1.48
Precision 6.17 4.12 0.47 0.39
Accuracy 96.09 107.04 100.93 99.66
Table 3 Systems suitability parameters for Acyclovir (ACV), n = 6.
System suitability parameter ACV
Standard working solution 400 ppm
Retention time 0.89 min
% RSD 0.47%
Capacity factor (k) 1.66
Theoretical plate number (N) 5069.4

Note:  The chromatographic conditions used were: Acquity® UPLC BEH C18 column (2.1 × 50 mm, 1.7 μm) kept at 50 °C, mobile phase was water with 0.25% formic acid (FA) delivered at a flow rate of 0.5 ml/min, a Photodiode Array (PDA) eλ detector at wavelength of 254 nm.

4.4.1.3
4.4.1.3 Precision

The results of intra-day and inter-day precision are presented in Table 1 and Table 2, respectively. The developed method was found to be precise as the intra-day standard deviation (SD) values (Table 1) of six replicate analyses were within the range of 0–1.87 ppm. In the analytical range of 1–400 ppm (conc.), the coefficient of variation (CV; precision) was only ranging from 0% to 4.49%. Moreover, the inter-day (Table 1) SD values of six replicate determinations in three consecutive days were between 0.14 and 2.84 ppm, whereas the CV being in the range of 0.47%–6.17%. These low values of both SD and CV during the intra-day and inter-day analysis indicate perfectly the precision of the current method (Karnes and March, 1993).

4.4.1.4
4.4.1.4 Specificity

The specificity of the developed UPLC method for ACV was investigated in order to obtain an indication of the possible interferences from the degradation product(s) if present. The result in Fig. 2D shows that there was no degradation products present in the ACV sample. It seems that ACV compound can be recovered completely from the lipid formulation (Fig. 2D). Therefore, the R value in this assay can be calculated based on the availability of ACV peak only (20). In addition, there were no significant interfering peaks present in randomly selected drug free lipid formulation (QCzero) samples at ACV retention time (Fig. 2C), which suggests that ACV compound can be analyzed predominantly from lipid based formulations.

4.4.1.5
4.4.1.5 Limit of Detection (LOD) and Lower Limit of Quantification (LLOQ)

The LLOQ in the assay was 1 ppm which was estimated to be the lowest concentration in the standard curve that can be measured with acceptable accuracy and precision for the ACV analyte, and with S/N ratio of 72.12. On the other hand, LOD was 0.1 ppm with S/N ratio of 14.15.

4.5

4.5 Application

Within the scope of the current project, the developed UPLC method has been proposed for the quantification of ACV compound in the studies of equilibrium solubility, dynamic dispersion and dissolution profiles of lipid-based formulations (Mohsin, 2012). A dispersion profile is shown in Fig. 5 as an example of the method application. The data represented an immediate release of ACV from a self-emulsifying lipid-based formulation in aqueous media (Mohsin and Alanazi, 2012). The lipid formulation consisted mixture of medium chain mono-glycerides (I308), water-soluble surfactant (T80) and cosolvent (Propylene glycol) with 5 mg of ACV solubilized in them. From the results it can be stated that the current method quantified more than 98% ACV, which was released immediately after dispersion and stayed in solution during 24 h time period.

The dispersion (immediate release profile) of ACV from a self-emulsifying lipid-based formulation in aqueous medium. Lipid formulation represents I308: PG (1:1)/T80 (1/1). Data are expressed as mean ± SD, n = 3.
Figure 5 The dispersion (immediate release profile) of ACV from a self-emulsifying lipid-based formulation in aqueous medium. Lipid formulation represents I308: PG (1:1)/T80 (1/1). Data are expressed as mean ± SD, n = 3.

5

5 Conclusion

The developed UPLC analytical method provides an ecofriendly, reliable, reproducible and specific assay for ACV in pure and pharmaceutical formulations. The described method is sensitive enough to detect as low as 0.1 ppm and exclusively offers a rapid determination of ACV (peak at 0.89 min within 1.2 min run time). No significant interferences were caused by the formulation excipients, diluents and or degradation products.

The validation method allows quantification of ACV in pure and pharmaceutical formulations particularly lipid based formulations in the range between 1 and 400 ppm. Compared to previously reported methods, the present assay method assessed extensive validation parameters as per FDA guidelines. The method has shown acceptable precision, accuracy and adequate sensitivity and demands to be in use for further studies.

The established method satisfies the system suitability criteria, peak integrity, and resolution of the drug peak. The overall results clearly indicate that the current method is attractive due to the good selectivity for quantitative determination of ACV in lipid-based formulation and also suitable for stability measurements.

Acknowledgments

Authors thankfully acknowledge the financial support provided by Kayyali Chair for Pharmaceutical Industries during this study. In addition, authors would like to thank Mr. Ahmed Shahba for his help during the UPLC method validation.

References

  1. , , , . Flow injection potentiometric determination of amantadine HCl. J. Pharm. Biomed. Anal.. 2002;30:601-611.
    [Google Scholar]
  2. , , , . Ultra performance liquid chromatography assay for cinnarizine in lipid-based formulations. Asian J. Chem.. 2012;24:595-600.
    [Google Scholar]
  3. , , , . Validated liquid chromatographic-fluorescence method for the quantitation of gemifloxacin in human plasma. Talanta. 2010;83:110-116.
    [Google Scholar]
  4. , , , . Stability indicating simultaneous determination of domperidone (DP), methylparaben (MP) and propylparaben by high performance liquid chromatography (HPLC) J. Pharm. Biomed. Anal.. 2006;41:358-365.
    [Google Scholar]
  5. , , , , , , , , , , , . Biowaiver monographs for immediate release solid oral dosage forms: aciclovir. J. Pharm. Sci.. 2008;97:5061-5073.
    [Google Scholar]
  6. , , , . Simple high-performance liquid chromatographic method for the determination of acyclovir in pharmaceuticals. Farmaco. 2003;58:1301-1306.
    [Google Scholar]
  7. , , , , . Development and validation of UPLC method for determination of primaquine phosphate and its impurities. J. Pharm. Biomed. Anal.. 2008;46:236-242.
    [Google Scholar]
  8. , , , , , , , . Precision and accuracy in the quantitative analysis of biological samples by accelerator mass spectrometry: application in microdose absolute bioavailability studies. Anal. Chem.. 2011;83:5607-5616.
    [Google Scholar]
  9. , , . Precision, accuracy, and data acceptance criteria in biopharmaceutical analysis. Pharm. Res.. 1993;10:1420-1426.
    [Google Scholar]
  10. , , , . Simple and sensitive spectrophotometric methods for determination of amantadine hydrochloride. J. Mass Spectrom.. 2002;37:477-480.
    [Google Scholar]
  11. , . Validation of HPLC methods in pharmaceutical analysis. In: , , eds. Separation Science and Technology. Academic Press; .
    [Google Scholar]
  12. , , , , , . Determination of acyclovir in blood serum and plasma by micellar liquid chromatography with fluorimetric detection. J. Liq. Chromatogr.. 1993;16:2359-2386.
    [Google Scholar]
  13. , . Design of lipid-based formulations for oral administration of poorly water-soluble drug fenofibrate: effects of digestion. AAPS PharmSciTech. 2012;13:637-646.
    [Google Scholar]
  14. , , . The fate of paclitaxel during in vitro dispersion testing of different lipid-based formulations. J. Drug Deliv. Sci. Technol.. 2012;22(2):197-204.
    [Google Scholar]
  15. , , . Linearity and the limitations of least squares calibration. J. Chromatogr. A. 1997;762:73-82.
    [Google Scholar]
  16. , , , . Advantages of application of UPLC in pharmaceutical analysis. Talanta. 2006;68:908-918.
    [Google Scholar]
  17. , , . Simple high-performance liquid chromatographic method for the determination of acyclovir in human plasma using fluorescence detection. J. Chromatogr. B Biomed. Sci. Appl.. 1997;693:241-244.
    [Google Scholar]
  18. , . Determination of amantadine in human plasma by capillary gas chromatography using electron-capture detection following derivatization with pentafluorobenzoyl chloride. J. Pharm. Biomed. Anal.. 1993;11:699-703.
    [Google Scholar]
  19. , , , , . Electrophoretic behavior of adamantane derivatives possessing antiviral activity and their determination by capillary zone electrophoresis with indirect detection. J. Electrophor.. 2002;23:259-262.
    [Google Scholar]
  20. , , . Scintillation proximity radioimmunoassay for the measurement of acyclovir. J. Pharm. Biomed. Anal.. 1996;15:157-163.
    [Google Scholar]
  21. , , , . Use of oxidation reactions for the spectrophotometric determination of acyclovir and amantadine hydrochloride in their dosage forms. J. Biomed. Chomatogr.. 2003;17:500-503.
    [Google Scholar]
  22. US DEPT. OF HEALTH AND HUMAN SERVICES, F. A. D. A., CENTRE FOR DRUG EVALUATION AND RESEARCH (CDER), CENTRE FOR VETERINARY MEDICINE (CVM) 2001. Guidance for Industry. Bioanalytical Method Validation. Rockville, MD, USA.
  23. , , , , . Development and validation of a plasma assay for acyclovir using high-performance capillary electrophoresis with sample stacking. J. Chromatogr. B. 2002;772:291-297.
    [Google Scholar]
  24. , , , . A simple high-performance liquid chromatographic method for the determination of acyclovir in human plasma and application to a pharmacokinetic study. Arzneimittelforschung. 2008;58:199-202.
    [Google Scholar]
Show Sections