On-line preconcentration of ultra-trace thallium(I) in water samples with titanium dioxide nanoparticles and determination by graphite furnace atomic absorption spectrometry
⁎Corresponding author. Address: Ferdowsi University Campus, Azadi Sq., Mashhad, Iran. Tel.: +98 511 8797022; fax: +98 511 8796416. s.asadpour@gmail.com (Saeid Asadpour)
-
Received: ,
Accepted: ,
This article was originally published by Elsevier and was migrated to Scientific Scholar after the change of Publisher.
Peer review under responsibility of King Saud University.
Abstract
A new method has been developed for the determination of Tl(I) based on simultaneous sorption and preconcentration with a microcolumn packed with TiO2 nanoparticle with a high specific surface area prepared by Sonochemical synthesis prior to its determination by graphite furnace atomic absorption spectrometry (GFAAS). The optimum experimental parameters for preconcentration of thallium, such as elution condition, pH, and sample volume and flow rate have been investigated. Tl(I) can be quantitatively retained by TiO2 nanoparticles at pH 9.0, then eluted completely with 1.0 mol L−1 HCl. The adsorption capacity of TiO2 nanoparticles for Tl(I) was found to be 25 mg g−1. Also detection limit, precision (RSD, n = 8) and enrichment factor for Tl(I) were 87 ng L−1, 6.4% and 100, respectively. The method has been applied for the determination of trace amounts of Tl(I) in some environmental water samples with satisfactory results.
Keywords
Preconcentration
Thallium(I)
Titanium dioxide nanoparticles
Microcolumn separation
Adsorption capacity
GFAAS
1 Introduction
Nowadays heavy metals such as thallium have received considerable attention owing to their effects in the pollution of the environment. Many thallium(I) compounds are highly soluble in water and are readily absorbed through the skin. It has been identified to be an environmentally important element due to its toxic effects and because it is commonly found with lead, zinc, iron, tellurium and the alkalis. Tl is known to have side effects such as mutagenic, carcinogenic and teratogenic. Thallium(I) is highly toxic to the biosphere even more than Hg, Cd, Pb and Cu as well (Leonard and Gerber, 1997; Nriagu, 1998). Thallium is used as a catalyst, in making alloys, optical lenses, low temperature thermometers, dyes, pigment in scintillation counters and is used as medicines, rodenticides and insecticides as well (Lan and Lin, 2005). In human beings, poisoning from toxicological exposure to thallium occur in case of homicide, suicide and inadvertence (Günther and Kastenholz, 2005).
Owing to the toxicity of this metal and low concentration of thallium in aqueous environment (Das et al., 2007), an analytical technique which is presenting high sensitivity and low detection limit is required. Regarding to their high selectivity and sensitivity, atomic absorption spectrometry and particularly graphite furnace atomic absorption spectrometry (GFAAS) methods are extensively used for determination of thallium in various samples (Manning and Slavin, 1988). However, direct determination of thallium by GFAAS includes many difficulties due to spectral and non-spectral interferences in environmental matrices (Schmidt and Dietl, 1983; Shan et al., 1984). Other alternative is a GFAAS determination of thallium followed by separation and preconcentration (Cvetkovic et al., 2002; Bundalevska et al., 2005).
Numerous analytical methods have been proposed for separation and preconcentration of thallium, mainly based on liquid–liquid extraction (Asami et al., 1996), single drop liquid-phase microextraction (SDME) (Chamsaz et al., 2009), solid-phase extraction (SPE) (Lin and Nriagu, 1999; Mobarakeh and Mahani, 2005; Dadfarnia et al., 2007), coprecipitation (Stafilov and Čundeva, 1998) and ion-exchange (Jain et al., 1980).
Of all above methods, solid-phase extraction has attracted great attention for the preconcentration of traces of heavy metal ions owing to its simple operation, rapid phase separation, no emulsification, higher preconcentration factor and easy automation and combination with different detection techniques (Pyrzynska, 1998; de Godoi Pereira and Arruda, 2003). The selection of solid-phase extractant is the decisive factor that affects analytical sensitivity and selectivity. Various adsorbents such as silica C-18 (Urbánková and Sommer, 2008), Chromosorb 105 resin (Karatepe et al., 2011), Amberlite XAD resins (KOSHIMA, 1986), polyethylene (do Nascimento and Schwedt, 1997), activated carbon (Koshima and Onishi, 1985), microcrystalline naphthalene (Rezaei et al., 2007) and carbon nanotubes (Pacheco et al., 2009) have been investigated for preconcentration of thallium using SPE based on adsorption.
The nanometer material is usually a new functional material (Liang et al., 2000), which has attracted much attention due to its special properties (Watzke and Fendler, 1987). Most of the atoms on the surface of the nanoparticles are unsaturated that easily bind with other atoms. Nanoparticles have attracted much attention due to their special properties and high capacity of sorption. Moreover, the procedure is not only simple, but also the sorption process is rapid. So there is a growing interest in the application of nanoparticles as sorbents (Maria Claesson and Philipse, 2007). Recently it has been reported that titanium dioxide nanoparticles have been successfully used for separation and preconcentration of trace metal ions (Yang et al., 2004; Liang et al., 2007).
Thus, this paper aims to focus on the application of TiO2 nanoparticles which are synthesized with a new method under ultrasonic irradiation at low intensity and to evaluate the adsorption properties of this nanomaterial for preconcentration of Tl(I) using column method and coupled with GFAAS for the determination of the metal in water samples. Statistical properties of the method developed are presented.
2 Experimental
2.1 Apparatus
All measurements were carried out using a Perkin Elmer AAnlyst 800 graphite furnace atomic absorption spectrometer (Perkin Elmer Instruments, Shelton, CT, USA). The pH values were controlled with a Mettler Toledo 320-S pH meter (Mettler Toledo Instruments Co. Ltd., Shanghai, China) supplied with a combined electrode. Four-channel minipuls3 peristaltic pump (Gilson, Inc., USA) was used in the preconcentration process. A self-made PTFE microcolumn (40 mm × 3.0 mm i.d.), packed with nanometer size TiO2, was used in all experiments. A minimum length of PVC tubing with an i.d. of 1.0 mm was used for all the connections.
2.2 Reagents and solutions
A stock solution (1000 mg L−1) was prepared by dissolving solid Thallium nitrite(I) in aqueous solution. Thallium nitrate(I) (microscopy grade) was purchased from (Acros Organics, Geel, Belgium) and was used without any further purification. Working standard solutions were obtained by appropriate dilution of the stock standard solutions. Buffer solution at pH 9 (Merck, Darmstadt Germany) was used for conditioning the column in the preconcentration processes. Titanium tetra-isopropoxide (TTIP) (Merck, Darmstadt, Germany), ethanol (analytical grade) and glacial acetic acid 100% (Merck, Darmstadt, Germany) were used without any further purification. Glass beads with 40–60 mesh, was purchased from Glass beads Company (Glass beads Co., Tehran, Iran).
2.3 Preparation of TiO2 nanoparticles
The titanium nanoparticles with diameter nearly 6 nm were prepared based on the method which is reported recently by Ghows and Entezari (2010). Fifty milliliters of deionized water and 0.2 mL acetic acid as a dispersant were mixed together in a sonication cell and sonicated. Then 2 mL titanium tetra-isopropoxide with 5 mL ethanol was injected drop-wise into an aqueous solution in 2 min. The mixture was sonicated continuously for 3 h at room temperature condition. The sonication was conducted without cooling so that the temperature was raised from 25 to 75 °C at the end of the reaction. The achieved precipitates were separated by centrifugation (20,000 rpm in 20 min) and washed three times with ethanol and deionized water. The product was dried at 40 °C for 24 h. Fig. 1a shows the TEM micrograph of TiO2 nanoparticles. As shown, the diameter of the particles is estimated to be less than 10 nm. HRTEM image shows that the sample is a single crystal with a growing direction and a step of one atomic layer can be seen as well (Fig. 1b). The crystallinity of the product is also proven by selected area electron diffraction (SAED) as a crystallographic experimental method (insert in Fig. 1b).TEM (a), HRTEM and SAED (b) images.
2.4 Column preparation
A homogenous mixture of 20 mg of TiO2 nanoparticles and 30 mg of glass beads with 40–60 mesh was introduced into a PTFE microcolumn (40 mm × 3.0 mm i.d.) plugged with a small portion of glass wool at both ends. Glass beads were used to facilitate the smooth flow of solution through the column and to probably increase the contact of solution with the surface of the nanoparticles. Before using, 1.0 mol L−1 HCl solution and doubly distilled deionized water were passed through the column in order to clean and condition it. Then, the column was conditioned to the desired pH with 10 mL of 0.1 mol L−1 NH4Cl buffer solution.
2.5 General procedure
A portion of aqueous sample solution containing Tl(I) was prepared and the pH value was adjusted to desired value with 0.1 M HCl and NaOH. The solution was passed through the column by using a peristaltic pump adjusted to the desired flow rate. As the preconcentration was completed, the retained metal ions on the microcolumn were eluted with 1.0 mL of 1.0 M HCl solution. The analytes in the effluents were determined by GFAAS. The column could be used repeatedly after regeneration with 1.0 M HCl or HNO3 solution followed by distilled water. Between the extractions, the minicolumn was dried by passing the air through it for 30 min.
3 Result and discussion
3.1 Effect of pH on adsorption
pH value plays an important role in the adsorption process due to the adsorption of different ions on TiO2 surfaces. As said by Morterra (1988), the pH of solution affects the distribution of active sites on the surface of titanium dioxide nanoparticles. The surface charge is neutral at the isoelectric point (IEP), which is at pH 6.8 for TiO2 nanoparticles (Zhang et al., 2009). In general, when the pH of the solution is higher than the isoelectric point of the adsorbent, the negative charge on the surface provides electrostatic attractions that are favorable for adsorbing cations such as thallium. In this experiment, the effect of the pH on the retention of thallium on the TiO2 was studied in the range of 4–12 and operated according to the recommended procedure.
Fig. 2 shows the results of the effect of pH on the recoveries of thallium(I) ions by nanoparticles, which indicates that the adsorption percentage of the thallium sharply increases to the maximum as the pH increases to 9 and keep constant with the further increase of pH. Hence, in the following experiments the sample pH was adjusted to 9.Effect of pH on the adsorption of Tl(I) on TiO2 nanoparticles; Tl(I): 1.5 μg L−1; sample volume: 50 mL.
3.2 Elution of the adsorbed Tl(I) ions
It is found from Fig. 2 that the adsorption of thallium at pH < 4 could be negligible. Thus, various concentration of HCl were considered for the elution of retained Tl(I) from the microcolumn. Fig. 3 shows the effect of HCl concentration on the recovery of the adsorbed Tl(I). It can be seen that 1.0 M HCl is enough to recover the Tl(I) quantitatively. Therefore, we applied 1.0 M HCl to recover the Tl(I) from microcolumn in this research. Then the effect of elution volume on the recovery of Tl(I) by keeping the HCl concentration equal to 1.0 M was studied. It was found that quantitative recoveries (>95%) could be achieved by 1.0 mL of 1.0 M HCl as the eluent.Effect of HCl concentration in the eluent on the recovery of Tl(I); Tl(I): 1.5 μg L−1; sample volume: 50 mL; pH: 9.0.
3.3 Effect of flow rate of sample solution
The flow rate of sample solution affects the retention of Tl(I) on the nanoparticle as an adsorbent and it controls the time of analysis. Thus the influence of flow rate of sample solution on retention of Tl(I) ions in a range of 0.5–2.5 mL min−1 under the optimum conditions (pH, eluent, etc.) was studied via passing 50 mL of sample solution through the microcolumn by a peristaltic pump. We observed that quantitative recovery (>95%) of Tl (I) was obtained up to a flow rate of 1.5 mL min−1 and the recovery decreased when the flow rate was greater than 1.5 mL min−1 due to a decrease in the adsorption kinetics of reaction at higher flow rate. Thus, a flow rate of 1.5 mL min−1 was employed for subsequent investigations.
3.4 Effect of sample volume
In order to investigate the possibility of achieving high levels of enrichment factor, we examined the effect of sample volume on recovery of Tl(I). To achieve to this goal, a series of sample solution over the range of 25–200 mL containing a fixed quantity of Tl(I) were processed based on the recommended procedure. The results were exhibited in Fig. 4.Effect of sample volume on the adsorption of Tl(I) on TiO2 nanoparticles; Tl(I): 1.5 μg L−1; pH: 9.0.
According to this figure, as shown, quantitative recovery (>95%) of thallium was obtained for sample volumes of ⩽100 mL for Tl(I) ions. Thus, the adsorbed Tl(I) can be eluted with 1.0 mL 1.0 mol L−1 HCl, so the enrichment factor of 100 was achieved by this method.
3.5 Adsorption capacity
Adsorption capacity is one of the important parameters in assessment of the adsorption capacity of different adsorbents, since it suggests how much adsorbent is required to quantitatively concentrate the analyte from a solution. The capacity study method which is used was the one adapted from that recommended by Maquieira et al. (1994). To determine the adsorption capacity, 30 mL aliquots of sample solutions containing the different Tl(I) concentrations (in the range of 0–30 μg mL−1) were adjusted to the appropriate pH with 0.1 mol L−1 HCl or NaOH. Then 20 mg of TiO2 nanoparticles was packed in microcolumn and the preconcentration procedure described above was applied. The amount of Tl(I) adsorbed at each concentration level was determined. The profile of the adsorption isotherm for thallium was gained by plotting the concentration (μg mL−1) of Tl(I) solution versus the milligrams of Tl(I) adsorbed per gram TiO2 nanoparticles, and shown in Fig. 5. The result from the adsorption isotherm shows that the adsorption capacity of TiO2 for Tl(I) was found to be 25 mg g−1.Adsorption isotherm of Tl(I) on TiO2 nanoparticles; pH: 9.0; sample volume 30 mL.
3.6 Column reuse
The stability and potential regeneration of the microcolumn packed with TiO2 nanoparticle were investigated. The column can be reused after regenerated with 10 mL 1.0 mol L−1 HCl followed by 20 mL deionized water. After repeating this process 15 times, the recovery of Tl(I) was still above 95%. Thus the microcolumn packed with TiO2 nanoparticles can be reused at least 15 times.
3.7 Effects of coexisting ions
The effect of common coexisting ions on the thallium retention onto TiO2 nanoparticles was evaluated using the developed method. The existence of coexistence ions would affect the preconcentration of thallium by competitive adsorption. In these experiments, various foreign cations and anions at various concentrations were added to 1.5 μg L−1 Tl(I) at pH 9 and enriched in microcolumn according to the recommended procedure. The tolerance limits of the coexisting ions, defined as the largest amount making the recovery of Tl(I) less than 95%, were given in Table 1. It can be seen that the presence of coexisting ions at a much higher level has no obvious effect on the adsorption of Tl(I) ion under the selected conditions.
Coexisting ions
Tolerance limit of ions (Wion/Wtl(I))
Na+, K+
1000
Ca2+, Mg2+
500
Fe3+, Al3+
200
Sb3+, Pb2+
50
Zn2+, Ni2+, Cu2+, Cd2+
200
PO43−, SO42−, Cl−, NO3−
500
3.8 Analytical performance of the method
The calibration curve for Tl(I), based on optimal conditions, was linear with a correlation coefficient of 0.9986 in the range of 0.6–2.5 μg L−1. Also in accordance with the definition of IUPAC (Long and Winefordner, 1983), the detection limit of this method, evaluated as the concentration corresponding to three times the standard deviation of 11 runs measurements of blank solution using the preconcentration method, was found to be 87 ng L−1 for Tl(I). The results (Table 2) show that the precision of this method (RSD), examined by eight replicate measurements of 1.5 μg L−1 Tl(I) in 100 mL of model solutions, was about 6.4% and indicating that the method has good precision for the analysis of trace Tl(I) ions in solution samples.
R2
Limits of detection (ng L−1)
Linear range (μg L−1)
RSD (%) (n: 8)
0.9986
87.0
0.6–2.5
6.4
Table 3 compares the characteristic data of the present method with some other methods reported in literatures. As can be seen, the enrichment factor and detection limits obtained by the present method are comparable to those reported method, and it can be used efficiently for sample containing thallium.
Adsorbent
Detection method
Species
Enrichment factor
Detection limit (μg L−1)
Reference
2,6-Bis(N-phenyl carbamoyl)pyridine naphthalene
UV–vis
Tl(III)
100
0.25
Rezaei et al., 2007
Chromosorb 105 resin
ETAAS
Total Tl
150
0.050
Karatepe et al. (2011)
Tl(III)
0.034
Thallium(III) ion imprinted polymer
GFAAS
Tl(III)
10
0.02
Arbab-Zavar et al. (2011)
Multiwalled carbon nanotubes
GFAAS
Tl(III)
20
0.009
Gil et al. (2009)
Immobilized oxine
FAAS
Tl (total)
100
2.5
Dadfarnia et al. (2007)
Thiol cotton fiber (TCF)
GFAAS
Tl(I)
–
0.003
Yu et al. (2001)
Sawdust
FAAS
Tl(I)
160
12.5
Memon et al. (2008)
Modified silica SGX C18
ICP-AES
Tl(III)
100
0.72
Urbánková and Sommer (2008)
Titanium dioxide nanoparticles
GFAAS
Tl(I)
100
0.087
This work
3.9 Application of the method on real waters
In order to establish the performance of preconcentration method in determination of Tl(I) in real samples, two environmental water samples (tap water and Mines lake water) were collected from the region of Serdang city in Malaysia and used for the mentioned analysis. In these experiments, the samples were spiked with 1.0 and 2.0 μg L−1 Tl(I) and subjected to the preconcentration procedure. The results (Table 4) indicate that proposed method could be used to determine thallium in real water samples.
Samples
Added
Found
Recovery (%)
Tap water
0
0.2 ± 0.2
1
1.1 ± 0.1
90.0
2
2.1 ± 0.09
95.0
Mines lake water
0
0.3 ± 0.2
1
1.2 ± 0.2
95.0
2
2.2 ± 0.2
96.5
The method has been applied for the determination of Tl(I) in the standard reference material (JR-1) for its thallium content as well. The rock sample (0.5 g) was dissolved in 10 mL of mixture of HF (25 mL), HNO3 (2 mL) and H2SO4 (8 mL) in a teflon beaker, and heated until 2 mL of solution was remained. Eight milliliters of concentrated HNO3 was added to this solution and after heat treatment distilled water was added to give a clear solution and was finally made to 100 mL by further addition of distilled water. This solution was analyzed for its Tl content according to the proposed procedure and was determined to be 1.55 ± 0.18 μg g−1 which is in good agreement with its certified value (1.56 μg g−1) with a recovery of 99.3%.
4 Conclusions
TiO2 nanoparticles have been successfully synthesized by Sonochemical method and used as the microcolumn packing material for preconcentration of Tl(I) prior to its determination by graphite furnace atomic absorption spectrometry. The results obtained indicate that TiO2 nanoparticles are effective SPE sorbents for the adsorption of Tl(I) from aqueous solutions. The precision and accuracy of proposed method were satisfactory and percentage of Tl(I) removal from solution is close to 100% at pH 9. Also The TiO2 nanoparticles prepared by this method show a remarkable adsorption capacity of 25 mg g−1 for Tl(I) ions. The method can be successfully applied to the separation, preconcentration and determination of Tl(I) in environmental samples.
References
- Synthesis and characterization of nano-pore thallium(III) ion-imprinted polymer as a new sorbent for separation and preconcentration of thallium. J. Hazard. Mater.. 2011;185(1):38-43.
- [Google Scholar]
- Determination of thallium in soils by flame atomic absorption spectrometry. Fresenius J. Anal. Chem.. 1996;356(5):348-351.
- [Google Scholar]
- Separation of tl(I) and tl(III) from environmental water samples by flotation method coupled with Zeeman Etaas determination. J. Environ. Sci. Health, Part A. 2005;40(5):1045-1054.
- [Google Scholar]
- Preconcentration of thallium(I) by single drop microextraction with electrothermal atomic absorption spectroscopy detection using dicyclohexano-18-crown-6 as extractant system. J. Hazard. Mater.. 2009;167(1–3):597-601.
- [Google Scholar]
- Determination of thallium in wine by electrothermal atomic absorption spectrometry after extraction preconcentration. Spectrochim. Acta, Part B. 2002;57(6):1101-1106.
- [Google Scholar]
- Speciation and determination of thallium by on-line microcolumn separation/preconcentration by flow injection-flame atomic absorption spectrometry using immobilized oxine as sorbent. J. Hazard. Mater.. 2007;148(1–2):446-452.
- [Google Scholar]
- Trends in preconcentration procedures for metal determination using atomic spectrometry techniques. Anal. Chim. Acta. 2003;141(3):115-131.
- [Google Scholar]
- Polyethylene powder as an absorbent for preconcentration of aluminium, beryllium and thallium. Microchim. Acta. 1997;126(1):159-166.
- [Google Scholar]
- Ultrasound with low intensity assisted the synthesis of nanocrystalline TiO2 without calcination. Ultrason. Sonochem.. 2010;17(5):878-883.
- [Google Scholar]
- Speciation analysis of thallium using electrothermal AAS following on-line pre-concentration in a microcolumn filled with multiwalled carbon nanotubes. Microchim. Acta. 2009;167(3):187-193.
- [Google Scholar]
- Günther, K., Kastenholz, B., 2005. Speciation of thallium. In: Cornelis, I.R. (Ed.), Handbook of Elemental Speciation II – Species in the Environment, Food, Medicine and Occupational Health. Wiley, New York, pp. 408–421.
- Application of a collidinium molybdoarsenate ion-exchange column to the separation of thallium(I) Analyst. 1980;105(1252):685-689.
- [Google Scholar]
- Selective preconcentration of thallium species as chloro and iodo complexes on chromosorb 105 resin prior to electrothermal atomic absorption spectrometry. Talanta. 2011;85:1974-1979.
- [Google Scholar]
- Adsorption of iron(III), gold(III), gallium(III), thallium(III) and antimony(V) on amberlite xad and chelex 100 resins from hydrochloric acid solution. Anal. Sci.. 1986;2(3):255-260.
- [Google Scholar]
- Adsorption of iron(III), gallium(III), thallium(III), antimony(V) and antimony(III) on activated carbon from hydrochloric acid solution (retroactive coverage) Anal. Sci.. 1985;1(3):237-240.
- [Google Scholar]
- Acute toxicity of trivalent thallium compounds to daphnia magna. Ecotoxicol. Environ. Saf.. 2005;61(3):432-435.
- [Google Scholar]
- Mutagenicity, carcinogenicity and teratogenicity of thallium compounds. Mutat. Res.. 1997;387(1):47-53.
- [Google Scholar]
- Study of the adsorption behavior of heavy metal ions on nanometer-size titanium dioxide with ICP-AES. Fresenius J. Anal. Chem.. 2000;368(6):638-640.
- [Google Scholar]
- Determination of trace rare earth elements by inductively coupled plasma optical emission spectrometry after preconcentration with immobilized nanometer titanium dioxide. Microchim. Acta. 2007;159(1):35-40.
- [Google Scholar]
- Thallium speciation in river waters with chelex-100 resin. Anal. Chim. Acta. 1999;395(3):301-307.
- [Google Scholar]
- Limit of detection. A closer look at the IUPAC definition. Anal. Chem.. 1983;55(7):712-724.
- [Google Scholar]
- The determination of thallium with the stabilized temperature platform furnace and Zeeman background correction. Spectrochim. Acta, Part B. 1988;43(9–11):1157-1165.
- [Google Scholar]
- Immobilized cyanobacteria for online trace metal enrichment by flow injection atomic absorption spectrometry. Anal. Chem.. 1994;66(21):3632-3638.
- [Google Scholar]
- Thiol-functionalized silica colloids, grains, and membranes for irreversible adsorption of metal (oxide) nanoparticles. Colloids Surf., A. 2007;297(1–3):46-54.
- [Google Scholar]
- Sawdust: a green and economical sorbent for thallium removal. Chem. Eng. J.. 2008;140(1–3):235-240.
- [Google Scholar]
- FAAS determination of thallium after preconcentration using nitroso-s and TDBA onto benzophenone. Bull. Chem. Soc. Ethiop.. 2005;19(1):17-22.
- [Google Scholar]
- An infrared spectroscopic study of anatase properties. Part 6.surface hydration and strong lewis acidity of pure and sulphate-doped preparations. J. Chem. Soc., Faraday Trans.. 1988;84(5):1617-1637.
- [Google Scholar]
- Nriagu, J.O., 1998. In: Nriagu, J.O. (Ed.), Advances in Environmental Science and Technology, Thallium in the Environment. Wiley, Chichester, p. 1.
- L-tyrosine immobilized on multiwalled carbon nanotubes: a new substrate for thallium separation and speciation using stabilized temperature platform furnace-electrothermal atomic absorption spectrometry. Anal. Chim. Acta. 2009;656(1–2):36-41.
- [Google Scholar]
- Recent advances in solid-phase extraction of platinum and palladium. Talanta. 1998;47(4):841-848.
- [Google Scholar]
- Preconcentration of thallium(III) with 2, 6-bis (N-phenyl carbamoyl) pyridine on microcrystalline naphthalene prior to its trace determination in human serum spectrophotometrically. Spectrochim. Acta Part A Mol. Biomol. Spectrosc.. 2007;67(1):92-97.
- [Google Scholar]
- Bestimmung von thallium in bodenaufschlüssen mit der flammenlosen atomabsorption in zirkonium-beschichteten graphitrohren. Fresenius J. Anal. Chem.. 1983;315(8):687-690.
- [Google Scholar]
- Application of matrix-modification in determination of thallium in waste water by graphite-furnace atomic-absorption spectrometry. Talanta. 1984;31(2):150-152.
- [Google Scholar]
- Determination of total thallium in fresh water by electrothermal atomic absorption spectrometry after colloid precipitate flotation. Talanta. 1998;46(6):1321-1328.
- [Google Scholar]
- The microdetermination of thallium by ICP-AES after previous preconcentration on modified silica. Comparison with integrated-platform graphite atomic absorption spectrometry. Anal. Chim. Acta. 2008;162(1):127-132.
- [Google Scholar]
- Quantum size effects of in situ generated colloidal cadmium sulfide particles in dioctadecyldimethylammonium chloride surfactant vesicles. J. Phys. Chem.. 1987;91(4):854-861.
- [Google Scholar]
- On-line separation and preconcentration of trace metals in biological samples using a microcolumn loaded with pan-modified nanometer-sized titanium dioxide, and their determination by ICP-AES. Microchim. Acta. 2004;144(4):227-231.
- [Google Scholar]
- Systematic studies on adsorption of 11 trace heavy metals on thiol cotton fiber. Anal. Chim. Acta. 2001;428(2):209-218.
- [Google Scholar]
- Sorption of thallium(III) ions from aqueous solutions using titanium dioxide nanoparticles. Microchim. Acta. 2009;165(1):73-78.
- [Google Scholar]